Maintenance and service
December 2022
12. If terminal block mounting
Install sample block heater rods
Figure 10-23: Installation of O
A. Cover
B. O-ring
C. Reference air tube
D. Heater rod
E. Sensor housing
F. Locking clip
G. Screw
H. Terminal insulator
I. Marking plate
J. Heater rod
K. Gasket
L. Screw
M. Heater strut assembly
Procedure
1. Before installing sample block heater rods, evenly coat the heater rods with
276
base mounting screws.
Watlube heater release agent.
(Figure
10-18) was moved, reinstall it with two
Cell and Heater Strut Assembly
2
Manual
00809-0500-4880