Configuration and start-up
December 2022
hardware to be properly installed external to the transmitter. A Rosemount
OCX8800 shipped from the factory must be configured before blowback can be
implemented. This same process must be performed any time a replacement card
stack is installed.
3.6.1 
Configure blowback with the Field Communicator -
HART
Procedure
1. Use the Field Communicator or AMS software to access the HART menu.
2. From the DETAILED SETUP menu, select INPUT/OUTPUT.
3. From the INPUT/OUTPUT menu, select BLOWBACK.
4. From the BLOWBACK menu, select BlBk Enabled. Select Yes to enable
5. Manually initiate blowback from DIAG/SERVICE.
3.6.2 
Configure blowback with the Field Communicator -
F
OUNDATION
Procedure
1. Use the Field Communicator or AMS software to access the Fieldbus menu.
2. From the TRANSDUCER block menu, select Alarm Relay/Blowback.
3. From the Alarm Relay/Blowback menu, select Blowback.
4. From the Blowback menu, select Blowback Enabled. Also set the following
60
®
blowback. Also set the following parameters:
BlBk Intrvl: Length of time between blowback events (60 minutes
recommended).
BlBk Period: Length of time blowback is activated (five seconds
recommended).
BlBk Purge Time: Length of time after blowback is complete before
oxygen / combustibles readings are considered valid (set as required by
the application).
Fieldbus
parameters:
Blowback Interval: Length of time between blowback events (60 minutes
recommended).
Blowback Period: Length of time blowback is activated (FIVE seconds
recommended).
Blowback Purge Time: Length of time after blowback is complete before
oxygen/combustibles readings are considered valid (set as required by
the application).
Initiate Blowback: Intiates a blowback event manually.