Emerson Rosemount OCX8800A Manual page 154

Oxygen and combustibles transmitter
Table of Contents

Advertisement

F
Fieldbus
OUNDATION
December 2022
Note
OUT = Block output value and status.
OUT_D = Discrete output that signals a selected alarm condition.
Figure 8-6: Analog Input Function Block Timing Diagram
A. OUT (mode in manual)
B. OUT (mode in automatic)
C. FIELD_VALUE
D. of change
E. Time (seconds)
8.8.3 
Filtering
The filtering feature changes transmitter's response time to smooth variations in
input readings caused by rapid changes in input.
You can adjust the filter time constant (in seconds) using the PV_FTIME
parameter. Set the filter time constant to zero to disable the filter feature.
154
Manual
00809-0500-4880
Emerson.com/Rosemount