Emerson Rosemount OCX8800A Manual page 62

Oxygen and combustibles transmitter
Table of Contents

Advertisement

Configuration and start-up
December 2022
4. From the CALIBRATION menu, select CAL VERIFY. Select Verify Calibration.
3.7.2 
Verify calibration with the Field Communicator -
FOUNDATION
Procedure
1. Use the Field Communicator or AMS software to access the Fieldbus menu.
2. From the TRANSDUCER block menu, select METHODS.
3. Set the Mode to OOS (out of service) before starting the calibration verify
4. From the METHODS menu, select OCX Cal Verify. From this menu, select the
3.7.3 
Verify calibration with the LOI
Procedure
1. Use the Z pattern to enter the LOI menu tree.
2. From the CALIBRATION menu, select Cal Verify.
3. From the Cal Verify menu, select the functions as follows:
62
From this menu, select the functions as follows:
Flow High O2 Gas: Flows the high O
calibration setup.
Flow Low O2 Gas: Flows the low O
calibration setup.
Flow High COe Gas: Flows the COe test gas for the time specified in the
calibration setup.
Purge Gas: Initiates a delay for the specified purge time before oxygen/
combustibles readings are considered valid.
Note
A purge automatically follows a gas flow.
Fieldbus
process.
functions as follows:
Flow High O2 Gas: Flows the high O
calibration setup.
Flow Low O2 Gas: Flows the low O
calibration setup.
Flow High COe Gas: Flows the COe test gas for the time specified in the
calibration setup.
Purge gas: Intiates a delay for the specified purge time before oxygen/
combustibles readings are considered valid.
Note
A purge automatically follows a gas flow.
test gas for the time specified in the
2
test gas for the time specified in the
2
test gas for the time specified in the
2
test gas for the time specified in the
2
Emerson.com/Rosemount
Manual
00809-0500-4880