Section 6: Troubleshooting
June 2012
Figure 6-1. Sensor module connector view
6.3.3

Removing the sensor from the housing

Note
To prevent damage to the senser module cable, disconnect it from the PlantWeb assembly
before you remove the senser module from the housing.
1.
2.
74
Wireless Housing
Loosen the housing rotation screw with a
turn.
Unscrew the housing from the senser module.
00809-0100-4100, Rev AA
3
/
-inch hex wrench, then rotate back one full
32