00809-0500-4880
Procedure
1. Apply pipe thread sealant (Loctite #567) to the exposed pipe threads of
2. Screw the sensor holder onto the eductor elbow.
3. With wrenches on eductor elbow and on flats of sensor holder, tighten the
4. Tighten the sensor holder to align outside flat with matchmark on sensor
Rosemount OCX8800A
eductor elbow
(Figure
Do not apply sealant to the first turn of the pipe threads.
sensor holder.
Do not allow the eductor elbow to turn.
housing flange, as shown in
Figure 10-21: COe Sensor Holder Alignment
A. Sensor holder flat
B. Matchmark
C. Straight edge
D. Matchmark
10-18).
Figure
10-21.
Maintenance and service
December 2022
273