N-560 Mode

Data Column Headings

Time

N-560 Operator's Manual
The last data field in the top line indicates the upper and lower alarm
limits for %SpO
and for the pulse rate (PR), and the SatSeconds alarm
2
setting (OFF). The SatSeconds setting may be OFF, 10, 25, 50, or 100
depending on the SatSeconds alarm setting. In the example above, the low
alarm limit for SpO
is 85% and the upper alarm limit is 100%. Pulse Rate
2
alarm limits are 40 to 170 bpm. SatSeconds alarm limit is set to off.
The N-560 mode is ADULT.
Actual column headings are in the third row of the column heading line.
Patient data that is presented in the chart, from left to right, are the time
that the patient data on the row was obtained, the current %SpO
being measured, the current Pulse Rate in beats per minute (bpm), the
current Pulse Amplitude (PA), and the operating status of the N-560.
The Time column represents the N-560 real-time clock. Refer to the
N-560 service manual to set the N-560 real-time clock.