Procedure

N-560 Operator's Manual
If the
button is pressed during the alarm silence duration,
Alarm Silence
the alarm silence duration is ended and the audible alarms are re-enabled.
Visual indications of an alarm condition cannot be turned off. For
example, if the %SpO
lower alarm limit is exceeded, the alarm can be
2
silenced for the alarm silence duration, but the %SpO
to flash.
If the alarm condition is still present when the alarm silence duration has
elapsed, the alarm will sound.
The power-on default setting for audible alarm silence duration is set at
the factory to 60 seconds. The default setting can be adjusted by service
personnel as described in the N-560 service manual.
With the N-560 in the normal monitoring mode:
1. Press the
Alarm Silence
are: OFF, 30, 60, 90, and 120 seconds.
Note: For steps 2 and 3, the
pressed while pressing the
2. While pressing the
Alarm Silence
button to increase alarm silence duration to Off, 30, 60,
Adjust Up
90, or 120 seconds.
button until XX SEC is displayed. Durations
and
Adjust Up
Adjust Down
button.
Alarm Silence
button, press and hold the
Using the N-560
value will continue
2
buttons must be
29