48
No hardware flow control is used. However, in the ASCII mode, XON/
XOFF flow control is supported.
TXD represents the Transmit Data line, and RXD is the Receive Data
line.
The pinouts for the data port are listed in Table 1.

Table 1: Data Port Pinouts

Pin
1
RXD+ (RS-422 positive input)
2
RXD (RS-232 input)
3
TXD (RS-232 output)
4
TXD+ (RS-422 positive output))
5
Signal Ground (isolated from Earth Ground)
6
NC (No connection)
7
Normally Open (relay closure nurse call, normally open)
8
Normally Closed (relay closure nurse call, normally closed)
9
RXD- (RS-422 negative output)
10
Signal Ground (isolated from earth ground)
11
Nurse Call (RS-232 level output)
12
TXD- (RS-422 negative output)
13
NC
14
NC
15
Nurse Call Common
The pin layouts (as viewed from the rear panel of the N-560) are
illustrated in Figure 3. The conductive shell is connected to earth ground