Printing; Printing N-560 Real-Time Data - Nellcor OXIMAX N-560 Operator's Manual

Hide thumbs Also See for OXIMAX N-560:
Table of Contents

Advertisement

P r i n t i n g

Printing N-560 Real-Time Data

N-560 Operator's Manual
Real-time data is continuously sent to the data port on the back of the
N-560. Patient data can be obtained through the data port by connecting
to a computer or serial printer. When a real-time printout or display is
being transmitted to a printer or computer, a new line is printed/displayed
every 2 seconds. Column headings are printed/displayed every 25 lines,
or if one of the values in the column heading is changed.
Note: If the data output stops transmitting, the N-560 must be turned off
and then turned back on.
When connecting the N-560 to a printer or computer, verify proper
operation before clinical use. Both the N-560 and the printer or PC must
be connected to a grounded AC outlet.
Any printer or computer connected to the N-560's data port must be
certified according to IEC Standard 60950. All combinations of
equipment must be in compliance with IEC Standard 60601-1-1 systems
requirements. Anyone who connects a printer or PC to the data output
port configures a medical system and is therefore responsible that the
system complies with the requirements of system standard IEC Standard
60601-1-1 and the electromagnetic compatibility system standard IEC
Standard 60601-1-2.
1. Data Port Connector

Printing

55