Procedure - Nellcor OXIMAX N-560 Operator's Manual

Hide thumbs Also See for OXIMAX N-560:
Table of Contents

Advertisement

Using the N-560

Procedure

22
1. Turn on the N-560 by pressing and holding the
for more than one second.
2. The N-560 displays/sounds:
Display
5 (in pulse rate left window)
6 (in pulse rate center window)
0 (in pulse rate right window)
n (in SpO
left window)
2
n (in SpO
center window)
2
3. The N-560 automatically starts the Power-On Self-Test (POST),
which tests N-560 circuitry and functions.
CAUTION: During POST (immediately after power-up), confirm
that all display segments and indicators light, and the speaker sounds
a 1-second pass tone.
4. While performing POST, the self-test display appears for
approximately 2 to 4 seconds. During this time:
button
Power On/Off
Sound
one beep tone
one beep tone
one beep tone
none
none

Advertisement