Table 8: Electromagnetic Emissions - Nellcor OXIMAX N-560 Operator's Manual

Hide thumbs Also See for OXIMAX N-560:
Table of Contents

Advertisement

N-560 Operator's Manual
Table 7: Cable Compliance (Cont.)
Cables Comply With: (Cont.)
• RF emissions, EN 55011, Class B/Group 1
• EN 60601-1-2; 2001
Cables and OxiMax Sensors (Cont.)
D-YS sensor
D-YSE sensor
D-YSPD sensor

Table 8: Electromagnetic Emissions

The N-560 is intended for use in the electromagnetic environment
specified below. The customer or the user of the N-560 should assure that it
is used in such an environment.
Emission Test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic
Class D
Emissions
IEC 61000-3-2
Voltage
Complies
fluctuations/flicker
emissions
IEC 61000-3-3
Specifications
Maximum
4 ft. (1.2 m)
4 ft. (1.2 m)
4 ft. (1.2 m)
Electromagnetic Environment —
Guidance
The N-560 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
The N-560 is suitable for use in
establishments, including diagnostic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Length
105