Obtaining Technical Assistance

80
The N-560 is designed for use in environments in which the pulse can be
obscured by electromagnetic interference. During such interference,
measurements may seem inappropriate or the N-560 may not seem to
operate correctly.
Disruption may be evidenced by erratic readings, cessation of operation,
or other incorrect functioning. If this occurs, the site of use should be
surveyed to determine the source of this disruption, and the following
actions taken to eliminate the source:
Turn equipment in the vicinity off and on to isolate the offending
equipment.
Reorient or relocate the interfering equipment.
Increase the separation between the interfering equipment and
this equipment.
The N-560 generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with these instructions, may cause
harmful interference with other devices in the vicinity.
If assistance is required, contact Nellcor's Technical Services
Department, 1.800.635.5267, or your local Nellcor representative.
For technical information and assistance, or to order parts or a service
manual, contact Nellcor's Technical Services Department,
1.800.635.5267, or your local Nellcor representative. The service manual
includes block diagrams and a parts list required by qualified personnel
when servicing the N-560.
When calling Nellcor's Technical Services Department, 1.800.635.5267,
or your local Nellcor representative, you may be asked to tell the
representative the software version number of your N-560.
The device version appears in the N-560 display each time the N-560
successfully completes the power-on self-test. Write the number down
and have it available whenever requesting technical assistance.