Table 9: Electromagnetic Immunity (Cont.)
The N-560 is intended for use in the electromagnetic environment specified
below. The customer or user of the N-560 should assure that it is used in such
an environment.
IEC 60601
Immunity Test
Test Level
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the AC mains voltage prior to application of the test
T
level.
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHZ to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Specifications
Compliance
Electromagnetic
Level
Environment — Guidance
Complies
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
3 Vrms
Portable and mobile RF
communications
equipment should be used
no closer to any part of the
3 V/m
N-560, including cables,
than the recommended
separation distance
calculated from the
equation applicable to the
frequency of the
transmitter.