Safety Notes And Protective Measures - Siemens POLYMOBIL III Service Instructions Manual

Hide thumbs Also See for POLYMOBIL III:
Table of Contents

Advertisement

1 - 2

Safety notes and protective measures

CAUTION
Connect the POLYMOBIL III only to a power source (power outlet) meeting the
standards of VDE 0107 or the usual local standards.
Before working on the open POLYMOBIL III switch OFF the unit with the power switch at
the control panele and unplug the power cable.
Before removing or inserting PC boards, switch off the generator; in doing so, observe
the ESD regulations.
POLYMOBIL III
When performing work and tests, the following must be
observed:
- the product-specific safety notes contained in the document,
- the safety notes RA0-000.012.40... in Register 2 of the
POLYMOBIL III file,
- and the general safety notes contained in Register 2 of the TI
binder.
After the system has been switched off, a dc voltage of approx.
300 V is applied for the main inverter.
On D920 the yellow LED V35 lights up.
Within approx. 1.5 minutes the voltage drops to approx. 12 V;
LED V35 goes out.
Checks and adjustments that must be carried out with radiation
switched on are marked with the symbol
When performing work procedures which are marked with this
symbol, radiation protective clothing must be worn.
RXB8-115.061.02
Rev. 01
Page 2 of 4
10.95
CS PS 24
Prerequisites
Siemens AG
Medical Solutions
1

Advertisement

Table of Contents
loading

Table of Contents