AC Voltage, Triangle wave Characteristics (typical)

0.3% of p-p value, from 10% to 90% point

AC Current (Sinewave) Extended Bandwidth Specifications

Ranges
All current ranges, <330 mA

AC Current (Non-Sinewave) Specifications

Trianglewave &
Truncated Sinewave
Ranges [1]
2.9 to 92.999 mA
93 to 929.999 mA
0.93 to 6.19999 A
6.2 to 31 A
[1] All waveforms are peak-to-peak output ranges.
[2] Uncertainty is stated in peak-to-peak. Amplitude is verified using an rms-responding DMM.
34
Linearity to 1 kHz
Frequency
0.01 - 10 Hz
10 - 10 kHz
Frequency
0.01 to 10 Hz
10 to 45 Hz
45 to 1 kHz
1 to 10 kHz
0.01 to 10 Hz
10 to 45 Hz
45 to 1 kHz
1 to 10kHz
10 to 45 Hz
45 to 1kHz
1 to 5 kHz
45 to 500 Hz
500 to 1 kHz
<1% of pk-to-pk value, with amplitude >50% of range
1-Year Absolute Uncertainty,
tcal + 5°C,
+ (% of output + % of range) [2]
%Output
5.0%
See "AC Current (Sinewave) Specifications"
1-Year Absolute Uncertainty,
tcal + 5°C,
+ (% of output + % of range) [2]
%Output
%Range
5.0%
0.5%
0.25%
0.5%
0.25%
0.25%
5.0%
0.5%
5.0%
0.5%
0.25%
0.5%
0.25%
0.5%
5.0%
1.0%
5.0%
1.0%
0.5%
0.5%
5.0%
1.0%
2.0%
0.5%
5.0%
1.0%
Aberrations
Maximum
Current
Resolution
%Range
0.5%
2 digits each
range
Maximum
Current
Resolution
Two digits, e.g., 75 mA
Six digits on each
range
Two digits
Six digits on each
range
Two digits
Six digits on each
range
Two digits on each
range
Six digits on each
range