4-42. Reading the UUT Error: Resistance Output

Proceed as follows to read the error of the UUT in the resistance function:
1. Set the desired resistance output as described under "Setting the Output."
2. Use the output adjustment controls as required to achieve a reading on the Control
Display (illustrated below) equal to the reading on the UUT. To adjust a higher-order
digit, press < . As you approach the reference value, work progressively back
towards the LSD (least-significant digit) on the calibrator's Control Display as
necessary by pressing > . You only need to go one digit past the LSD of the UUT;
digits below that are beyond the resolution of the UUT.