Illumination area and Measurement areas

The CM-3700A-U allows you to select an illumination area according to the specimen and your application.

 Illumination area


The CM-3700A-U allows you to select a SAV target mask (5 x 7 mm) or USAV (for 3 x 5 mm), according to
the specimen and your application.
The target mask surface which faces into the integrating sphere has a high-reflectance white coating. Since
the reflectance of this coating affects measurements, be careful to protect the white surface from scratches
and never touch the white surface.

 Measurement area


The measurement area is fixed to one kind: 1 x 3 mm measurements.