To Run The Self-Tests - HP 1660CS-Series User Manual

Logic analyzers
Table of Contents

Advertisement

To run the self-tests

Self-tests identify the correct operation of major functional areas of the
instrument. You can run all self-tests without accessing the interior of the
instrument. If a self-test fails, the troubleshooting flowcharts instruct you to
change a part of the instrument.
These procedures assume the files on the PV disk have been copied to the
/SYSTEM subdirectory on the hard disk drive. If they have not already been
copied, insert the PV disk in the flexible disk drive before starting this
procedure.
If you just did the power-up self-tests, go to step 2.
1
If you did not just do the power-up self-tests, disconnect all inputs,
then turn on the power switch. Wait until the power-up tests are
complete.
Press the System key, then select the field next to System. Then,
2
select Test in the pop-up menu.
Select the box labeled Load Test System, then select Continue.
3
Troubleshooting
To run the self-tests
12-15

Hide quick links: