Time Base - HP 1660CS-Series User Manual

Logic analyzers
Table of Contents

Advertisement

Menus and fields changed by the Autoscale algorithm
The following table shows the menus and their fields that are changed by the
autoscale algorithm.
Settings Changed by Autoscale
Menu
Field
Channel
V/Div
Offset
Trigger
Mode
Source
Level
Slope
Count
Auto-Trig
All Applicable
s/Div
Delay

Time base

The s/Div and Delay fields are displayed on all of the oscilloscope menus,
except for the Calibration menu.
s/Div field
The s/Div field allows you to set the sweep speed (time scale) on the
horizontal axis of the display from 1 ns/div to 5 s/div. Sweep speed is
measured in seconds per division.
Delay field
The Delay field allows you to set the horizontal position of the displayed
waveform in relation to the trigger. Delay time is always measured from the
trigger point on the waveform to the center of the screen. Delay time is
measured in seconds. When acquisitions are stopped, the Delay field can be
used to control the portion of acquisition memory displayed on screen.
Autoscale Action
Scaled - depending on amplitude of input signal.
Scaled - depending on offset of input signal.
Defaults to Edge.
Set to lowest numbered channel with signal present.
Scaled - depending on amplitude of lowest numbered channel with signal present.
Defaults to Positive.
Defaults to 1.
Defaults to On.
Scaled - depending on frequency of lowest numbered channel with signal present.
Defaults to 0 s.
Oscilloscope Common Menus
Time base
7-73

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

This manual is also suitable for:

1662cs1663cs1660cs1661cs

Table of Contents