To Run The Self-Tests - HP 1660 Series Service Manual

Logic analyzers
Hide thumbs Also See for 1660 Series:
Table of Contents

Advertisement

Troubleshooting

To run the self-tests

To run the self-tests

Self-tests identify the correct operation of major functional areas of the instrument. You can
run all self-tests without accessing the interior of the instrument. If a self-test fails, the
troubleshooting flowcharts instruct you to change a part of the instrument.
If you just did the power-up self-tests, go to step 2.
1 1
If you did not just do the power-up self-tests, disconnect all inputs, then turn on the
power switch. Wait until the power-up tests are complete.
Press the System key, then select the field next to System. Then, select Test in the
2 2
pop-up menu.
Insert the disk containing the performance verification tests (self-tests) into the disk
3 3
drive. Select the box labeled Load Test System, then select Continue.
Press the System key, then select then field next to Sys PV. Select System Test to
4 4
access the system tests.