Error Messages

This section lists some of the messages that the analyzer displays
when it encounters a problem.

". . . Inverse Assembler Not Found"

This error occurs if you rename or delete the inverse assembler file that is
attached to the configuration file. Ensure that the inverse assembler file is
not renamed or deleted, and that it is on the same flexible disk or in the same
directory as the configuration file.

"No Configuration File Loaded"

This is usually caused by trying to load a configuration file for one type of
module or the system into a different type of module.
Verify that the appropriate module has been selected as the target of the
Load operation. Selecting Load All will cause incorrect operation when
loading most preprocessor interface configuration files.
See Also
"To Load a Measurement Configuration" in chapter 6, "File Management."

"Selected File is Incompatible"

This occurs when you try to load a configuration file for the wrong module.
Ensure that you are loading a translatable configuration file for your logic
analyzer.