To arm the oscilloscope with the analyzer

If both analyzers and the oscilloscope are turned on, you can configure one
analyzer to arm the other analyzer and the oscilloscope. An example of this
is when a state analyzer triggers on a bit pattern, then arms a timing analyzer
and the oscilloscope which capture and display the waveform after they
trigger.
In the Analyzer Trigger menu, select Arming Control.
1
Select the Analyzer Arm Out Sent From field, and choose from the
2
list the Analyzer that will generate the arm.
Select the Analyzer Arm In field, and choose Group Run.
3
This allows you to time-correlate the data from the analyzers and the scope.
The Scope Trigger Mode must be Immediate for correlation.
Select the field of the instrument which will arm the others, and in
4
the pop-up set it to run from Group Run.
The Scope field is not selectable. To set how the scope is run, select the field
under Scope Arm In.
Select the other instrument fields and choose the mechanism which
5
will arm them.
The Analyzer Arm Out field determines which analyzer sends the arming
signal to Port Out and to the oscilloscope if the oscilloscope is being armed
by an analyzer.
As you set each machine, the arrows connecting the fields in the Arming
Control menu change. The arrows show the order in which the instruments
are armed.
See the example on the next page.
Select Done until you are back at the Trigger menu.
6
Using the Trigger Menu
To arm the oscilloscope with the analyzer
4–13