Configuring Volume Flow Measurement For Gas - Emerson Micro-Motion 2400S Configuration And Use Manual

Transmitters with analog outputs
Table of Contents

Advertisement

Optional Configuration
Table 8-1
Configuration map continued
Topic
Subtopic
Digital
Modbus address
communication
Modbus ASCII support
settings
HART address
Loop current mode
Infrared port write-protect
Floating-point byte order
Additional communications
response delay
Digital fault indicator
Burst mode
PV, SV, TV, QV assignments
Device settings
Sensor parameters
(1) These parameters apply only to transmitters with a display.
8.2

Configuring volume flow measurement for gas

Note: Volume flow measurement for gases cannot be configured using the Communicator. If the
flowmeter is configured to use a gas standard volume flow unit, the Communicator will display the
correct value, but will display "Unknown Enumerator" for the units label.
The display cannot be used to configure Vol Flow Type. However, after Vol Flow Type has been
configured using ProLink II, the volume flow unit can be configured.
Special functionality is provided in ProLink II for measuring the volume flow of gases. To access this
functionality:
1. Click
ProLink > Configure > Flow
2. Set
Vol Flow Type
3. Select the measurement unit you want to use from the
The default is
Note: When Vol Flow Type is set to Std Gas Volume, this list contains the units that are most
frequently used for gas measurement. If Liquid Volume is configured, gas measurement units are not
available.
4. Configure the
5. If you know the standard density of the gas that you are measuring, enter it in the
Density
following section.
Note: The term "standard density" refers to the density of the gas at reference conditions.
Note: Ensure that the values entered here are correct, and that fluid composition is stable. If either of
these conditions is not met, gas flow measurement accuracy will be degraded.
70
to
Std Gas Volume
.
SCFM
Std Gas Vol Flow Cutoff
field. If you do not know the standard density, you can use the Gas Wizard. See the
Tool
ProLink II
Communicator
✓ (partial)
.
.
Std Gas Vol Flow Units
(see Section 8.4). The default is
®
Micro Motion
Model 2400S Transmitters with Analog Outputs
Display
Section
8.11.1
8.11.2
8.11.3
8.11.4
8.11.5
8.11.6
8.11.7
8.12
8.13
dropdown list.