Inova SV 125 Operating Manual page 206

Fully automatic filling and closing machine for syringes
Table of Contents

Advertisement

5.2.54
Format input - - Stopper setting vacuum [3/3]
Touch control field
[2/3].
The parameters for vacuum stopper insertion can be entered.
Vacuum:
Vacuum stop [ms]
When the entered time has passed, the evacuation process during vacuum stop-
per insertion is aborted.
Vacuum start stopper setting [ms]
When the entered time has passed, and the pressure that has been entered on
field "Vacuum start stopper setting [mbar]" has been reached, vacuum stopper in-
sertion starts.
Vacuum start stopper setting [mbar]
When the entered pressure has been reached and the time that has been set on
field "Vacuum start stopper setting [ms]" has passed, vacuum stopper insertion
starts.
Version 1.0
5--120
on screen Format input - - Stopper setting vacuum
Operation
SV 125
4016589