On the teach--in screen for the needle stroke the needle is moved to the step points.
On the Format input screen the detected values are entered.
NOTE
Watch travel movement of axis to avoid collision.
Mode of procedure
Select input field "Start dosage position" on screen 691 and enter the value
1
for the filling start (in this example 190_).
Touch input field "No. of step points" and enter the desired number of step
2
points (in this example 3).
Call up screen Format input - - Teach servo axis needle stroke and touch
3
control field
inge. Note down the value.
Touch control field
4
field
move the filling needle approx. 2mm above fill level. Note down the value. For
the two remaining step points touch control field
field
To end the process move the filling needle out of the syringe to its upper posi-
5
tion above the syringe. Note down the value.
Version 1.0
5--96
to move the filling needle to the lowest position in the syr-
. The dosing is performed in steps. Touch control
and move the needle to the first step point. We recommend to
to move the axis to the step points. Note down the values.
Operation
SV 125
4016589
and touch control