Inova SV 125 Operating Manual page 178

Fully automatic filling and closing machine for syringes
Table of Contents

Advertisement

Delay start upper position [ms]
Delay of the start of the needle movement to upper position.
Start dosage position [inc]
Position of the rotary piston pumps when filling starts.
The position must lie between upper and lower position.
End dosage position [inc]
Position of the rotary piston pumps when filling ends.
180°
180°
Dosing start
No. of step points
Enter the number of step points. If the number of step points is reduced, all addi-
tional step points are deleted. The needle stroke does not move to the deleted step
points any longer.
Step point no.
Selection of a step point.
Step point position [mm]
Entry of the step point position of the selected step point number.
Segment type
Selection of a filling curve for a step point. To each step point a segment type can
be assigned. Either a linear or cubic filling curve can be selected (see sketch be-
Version 1.0
5--92
Dosage
320°
250°
The range for the dosing start is between
180° to 250°.
The dosing range is determined
within the area of 180° to 320°.
Operation
SV 125
4016589