Recommended Test Equipment - Agilent Technologies 35670A Installation Manual

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Agilent 35670A

Recommended Test Equipment

The following table lists the recommended equipment needed to test the performance
of the Agilent 35670A Dynamic Signal Analyzer. The table on page 1-20 lists
additional equipment needed to adjust and troubleshoot the analyzer. Other equipment
may be substituted for the recommended model if it meets or exceeds the listed critical
specifications. When substitutions are made, you may have to modify the procedures
to accommodate the different operating characteristics.
Recommended Test Equipment
Instrument
Critical Specifications
AC Calibrator
10 Hz to 102.4 kHz; 1 mV to 10 V
Amplitude
Amplitude Accuracy: ±0.1% phase locking
capability
Frequency
Frequency Range: 10 Hz to 1 MHz
Frequency Accuracy: ≤5 ppm
Synthesizer
Amplitude Accuracy:
0.2 dB from 1 Hz to 100 kHz
1 dB from 100 kHz to 1 MHz
Harmonic Distortion: ≤–70 dBc
Spurious: ≤–70 dBc
<±1 deg phase shift between output and
sync
Low Distortion
Frequency Range: 10 Hz to 100 kHz
Harmonic Distortion: ≤–93 dB, 10 Hz to
Oscillator
20 kHz
Digital
5 1/2 digit True rms ac Voltage:
Multimeter
30 Hz to 100 kHz; 0.1 to 500 V; ±0.1%;
≥1 MΩ input impedance dc Voltage:
1 V to 300 V; ±0.1%
50 Ω: ±2% at dc
Feedthrough
Termination (2)
(4 for option
AY6)
† John Fluke Manufacturing Co., Inc., PO Box C9090, Everett, WA 98206 U.S.A. (206) 347-6100
‡ Wavetek, 5808 Churchman Bypass, Indianapolis, IN 46203 U.S.A.
†† This equipment is not required for Operation Verification. The parts and schematic for the notch
filter are shown on page 1-19.
‡‡ ITT Pomona Electronics, 1500 East Ninth Street, Pomona, CA 91769 U.S.A. (714) 469-2900
FAX (206) 629-3317
Specifications
Recommended Test Equipment
Recommended Model
Fluke 5700A †
Alternate
Fluke 5200A †
Datron 4200, 4700, or 4708
HP 745A
HP 3326A
Alternate
(2) HP 3325A/B Opt 001
HP 339A ††
Alternate
HP 3326A with notch filter
††
HP 3325A/B with notch
filter††
HP 3458A
Alternate
HP 3456A , HP 3455A
HP 3478A
Pomona Elect Model 4119-50
‡‡
Alternate
HP 11048C, HP 10100C