To Run The Program In Manual Mode - Agilent Technologies 35670A Installation Manual

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Verifying Specifications

To run the program in manual mode

To run the program in manual mode
Use this procedure if you want to run the program in manual mode. You will be
prompted to set up all test equipment and you can check the analyzer's setup state after
each measurement.
1
Write in the information needed on the title page of the selected test record.
The test records are located near the back of this chapter and may be copied without
written permission of Agilent Technologies.
2
Press the following keys and when the program prompts you, set all GPIB addresses to
0:
[
[
[
[
[
[
[
3
Press the following keys:
[
[
[
[
[
[
4
Press the following keys to start the test:
[
[
5
Now follow the directions on the display and record the measurement result in the
selected test record after every measurement.
If you want to view the analyzer's setup state, press [
[
[
any instrument setup states, press [
ensure accurate measurement results.
Warning
During the test, the program prompts you to change the test equipment
connections. Always turn the ac calibrator output to OFF or STANDBY before
changing test equipment connections. The ac calibrator can produce output
voltages that could result in injury or death to personnel.
The directions on the display briefly tell you how to connect test equipment. For
detailed illustrations of equipment setup, see the setup illustrations starting on the next
page.
3-12
]
EQUIP CONFIG
]
AC CALIBRATO
]
SYNTH. 1
] (If needed)
SYNTH. 2
] (If needed)
LOW-D OSCILLATO
]
MULTIMETER
]
RETURN
]
TEST CONFIG
]
PROCEDURE
] or [
OP_VERIFY
PERFORMAN
]
STOP AFTER
EACH MEASUREMENT
]
RETURN
]
START TESTING
]
START BEGINNING
] or [
MEASURMNT STATE
] [
BASIC
DISPLAY SETUP
]
]
]. To continue the program, press
INPUT STATE
] [
] [
LOWER
RETURN
RESTART TEST
Agilent 35670A
] [
BASIC
Disp Format
] [
]. If you changed
CONTINUE
] instead of [
CONTINUE
]
] to

Advertisement

Table of Contents
loading

Table of Contents