Agilent 35670A
ITM_35670A Main Menu Descriptions
If you do not have a keyboard
connected to the analyzer, use the
numeric key pad and the alpha keys to
enter names or numbers. See the
analyzer's help text for a description of
the alpha keys.
Load and run the ITM_35670A
program to display the following
softkeys:
[ START TESTING ]
Displays a menu that allows you to start
testing with any test or to select just one
test in the list. Before pressing this
softkey, use [ TEST CONFIG ] and
[ EQUIP CONFIG ].
[ TEST CONFIG ]
Verifying Specifications
Displays the test configuration and a
menu that allows you to enter the
procedure, stop conditions, beeper
prompt, and GPIB address for the
analyzer and printer.
[ EQUIP CONFIG ]
Displays the test equipment
configuration and a menu that allows
you to enter the model number,
calibration due date, serial number, and
GPIB address for each test instrument.
[ TITLE PAGE ]
Displays the test record title page
information and a menu that allows you
to enter information for the analyzer.
[ STOP ITM ]
Stops the ITM_35670A program.
3-49