Preparing the Analyzer for Use
To change the fuses
To change the fuses
Both fuses are installed at the factory.
1
Unplug the power cord from the analyzer.
2
Press in and turn the appropriate fuse holder cap counter-clockwise (use a small screw
driver for the ac fuse). Remove when the fuse cap is free from the housing.
3
Pull the fuse from the fuse holder cap.
4
To reinstall, select the proper fuse and place in the fuse holder cap.
DC Fuse
HP 2110-0920 30 A 32 V Normal Blow
5
Place the fuse holder cap in the housing. Press in and turn clockwise.
2-10
Agilent 35670A
AC Fuse
HP 2110-0342 8 A 250 V Normal Blow