Recommended test equipment

Recommended test equipment

The following table is a list of the test equipment required to test performance, calibrate and
adjust, and troubleshoot this instrument. The table indicates the critical specification of the test
equipment and for which procedure the equipment is necessary. Equipment other than the
recommended model may be used if it satisfies the critical specification listed in the table.
Recommended Test Equipment
Equipment Required
Critical Specifications
Signal Generator
1 - 2.25 GHz, sine wave, amplitude 30 - 200
mV
RF Amplifier
22 dB gain at 10 MHz, 1.3 GHz bandwidth
1 - 500 MHz, -70 dBm to +44 dBm, ±3% accuracy
Power Meter/Power
Sensor
DMM
6 1/2 digit (0.1 mV) resolution, dcV accuracy
8 ppm/year, 4-wire resistance acc. ±0.25%
Power Supply
7 mV - 30 V dc, 0.1 mV accuracy and resolution
50 Ω type N, outputs differ by <0.15 dB
Power Splitter
Probe
No substitute
Probe Tip Adapter
1160 series to BNC
Oscilloscope
General-purpose
0.18 µF
Blocking Capacitor
Cable
Type N (m)(m) - 3 foot
Cable (2)
BNC - 3 foot
Cable (3)
BNC - 9 inch
Adapter
N (m) to BNC (m)
Adapter
N (m) to BNC (f)
Adapter
N (f) to BNC (m)
Adapter (2)
BNC tee (m)(f)(f)
Adapter
BNC (f)(f)
Adapter (2)
BNC (f) to dual banana (m)
BNC connectors 50 Ω
Termination
Shorting cap
BNC
5 Ω, 5 W
Resistor (2)
Video monitor
Accepts VGA-standard video signals
Keyboard
PC-compatible, AT 5-pin connector
Mouse
PS/2 compatible
POST Card
Power-on self test card compatible with PC-
compatible systems
Fan Safety Shield
Clips onto side of chassis with cover removed
Service software
No substitution
Note: the Fan Safety Shield and Service software are part of the Service Kit, Agilent P/N 54810-68703.
* P = Performance Tests, A = Adjustments, T = Troubleshooting
1–16
, time base accuracy 0.25 ppm
rms
Recommended Model
Use *
Agilent 8664A
P
Agilent 8447D/F
P
Agilent EPM-
P
441A/Agilent 8482A
Agilent 34401A
P, A, T
Agilent 6114A
P
Agilent 11667A
P
Agilent 1161A
P
Agilent 5063-2143
P
Agilent 54622A
P, T
Agilent 10240B
P
Agilent 11500A or B
P
Agilent 10503A
P, A, T
Agilent 10502A
P, A, T
Agilent 1250-0082
P
Agilent 1250-0780
P
Agilent 1250-0077
A
Agilent 1250-0781
P, T
Agilent 1250-0080
T
Agilent 1251-2277
P
Agilent 10100C
P
Agilent 1250-0774
P
Agilent 0812-0047
T
T
Agilent E2610A
T
Agilent C3751-60201
P, T
T
Agilent 54801-00601
A, T
Agilent 54810-68700
T