To Load The Program - Agilent Technologies 35670A Installation Manual

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Agilent 35670A

To load the program

For information about the program's softkeys, see the menu descriptions starting on
page 3-49.
1
Set the Agilent 35670A Dynamic Signal Analyzer's power switch to off (
connect the analyzer, test instruments, and printer using GPIB cables.
2
If you have the PC Style Keyboard, option 1CL, connect the keyboard to the analyzer
using the keyboard cable (see ''To connect the optional keyboard'' in chapter 2).
3
Insert the Semiautomated Performance Test Disk into the analyzer's disk drive, then
set the power switch to on (
4
After the analyzer finishes its power-up calibration routine, press the following keys:
[
]
Local/GPIB
[
SYSTEM CONTROLLR
[
]
System Utility
[
MEMORY USAGE
[
REMOVE WATERFALL
[
CONFIRM REMOVE
[
]
RETURN
[
]
MORE
[
]
SERVICE TESTS
[
PERFRMANC TEST
5
Now go to one of the following procedures to continue:
''To run the program in semiautomated mode''
''To run the program without a printer''
''To run the program in manual mode''
l
).
]
]
]
]
]
Verifying Specifications
To load the program
O
), then
3-7