Agilent 35670A
Test Configuration Menu Descriptions
Press [ TEST CONFIG ] to display the
test configuration and the following
softkeys:
[ Agilent 35670A ADDRESS ]
Prompts you to enter the GPIB address
for the Agilent 35670A Dynamic Signal
Analyzer.
The GPIB addresses equals 100 ö
(interface select code) + (primary
address). The interface select code for
the printer and test equipment is 7 (for
example, if the primary address is 8, the
GPIB address is 708).
[ PRINTER ADDRESS ]
Prompts you to enter the GPIB address
for the printer. To disable the printer,
set the printer address to 0.
[ PROCEDURE ]
Prompts you to select the operation
verification procedure (OP_VERIFY)
or the performance test procedure
(PERFORMAN).
Verifying Specifications
[ BEEPER ]
Toggles the beeper on and off. When
the beeper is on, the program beeps
approximately every 2 minutes while
waiting for you to follow the directions
on the display and press [ CONTINUE ].
[ RETURN ]
Returns to the ITM_35670A main menu.
[ STOP AFTER ]
Prompts you to select stop after limit
failure, stop after each measurement, or
do not stop after a limit failure or
measurement. If [ Limit Failure ] is
selected, the program stops after the
failing measurement is displayed but
before it is printed. At this point you
can continue on and print the failing
measurement or restart the
measurement.