Agilent Technologies 35670A Installation Manual page 151

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Agilent 35670A
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
Spurious Signals (continued)
Measurement
Two Ch, 32000 Hz Start, Ch 1
Two Ch, 32000 Hz Start, Ch 2
Two Ch, 33600 Hz Start, Ch 1
Two Ch, 33600 Hz Start, Ch 2
Two Ch, 35200 Hz Start, Ch 1
Two Ch, 35200 Hz Start, Ch 2
Two Ch, 36800 Hz Start, Ch 1
Two Ch, 36800 Hz Start, Ch 2
Two Ch, 38400 Hz Start, Ch 1
Two Ch, 38400 Hz Start, Ch 2
Two Ch, 40000 Hz Start, Ch 1
Two Ch, 40000 Hz Start, Ch 2
Two Ch, 41600 Hz Start, Ch 1
Two Ch, 41600 Hz Start, Ch 2
Two Ch, 43200 Hz Start, Ch 1
Two Ch, 43200 Hz Start, Ch 2
Two Ch, 44800 Hz Start, Ch 1
Two Ch, 44800 Hz Start, Ch 2
Two Ch, 46400 Hz Start, Ch 1
Two Ch, 46400 Hz Start, Ch 2
Two Ch, 48000 Hz Start, Ch 1
Two Ch, 48000 Hz Start, Ch 2
Two Ch, 49600 Hz Start, Ch 1
Two Ch, 49600 Hz Start, Ch 2
One Ch, 79200 Start, Ch 1
One Ch, 80800 Start, Ch 1
One Ch, 85600 Start, Ch 1
Operation Verification Test Record - Two Channel
Lower
Upper Limit
Limit
(dBfs)
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
5 of 10
Verifying Specifications
Measured Value
Pass/Fail
(dBfs)

Advertisement