Verifying Specifications

Measurement Uncertainty

Measurement Uncertainty
The following table lists the measurement uncertainty and ratio for each performance
test using the recommended test equipment. Except for the External Trigger test, the
ratios listed for the recommended test equipment meet or exceed the measurement
uncertainty ratio required by U.S. MIL-STD-45662A.
If you are using equipment other than the recommended test equipment, you may
calculate and record the measurement uncertainty and ratio for each performance test.
The table may be reproduced without written permission of Agilent.
Performance Test
Self Test
DC Offset
Noise
Spurious Signals
Amplitude Accuracy
-51 dBVrms
-43 dBVrms
-35 dBVrms
-27 dBVrms
-11 dBVrms
1 dBVrms
9 dBVrms
19 dBVrms
27 dBVrms
Flatness
25.6 kHz, 27 dBVrms
25.6 kHz, 9 dBVrms
25.6 kHz, -11 dBVrms
51.2 kHz, 27 dBVrms
51.2 kHz, 9 dBVrms
51.2 kHz, -11 dBVrms
99.84 kHz, 27 dBVrms
99.84 kHz, 9 dBVrms
99.84 kHz, -11 dBVrms
NA (not applicable) internal test
3-54
Using Recommended Test Equipment
Measurement Uncertainty
NA
NA
NA
NA
±0.020 dB
±0.0084 dB
±0.004 dB
±0.003 dB
±0.001 dB
±0.0008 dB
±0.001 dB
±0.00081 dB
±0.00117 dB
±0.01487 dB
±0.01277 dB
±0.01277 dB
±0.02025 dB
±0.01460 dB
±0.01583 dB
±0.02025 dB
±0.01460 dB
±0.01583 dB
Using Other Test Equipment
Ratio
Measurement Uncertainty
NA
NA
NA
NA
NA
NA
NA
NA
7.7:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
>10:1
10:1
>10:1
>10:1
10:1
>10:1
>10:1
Agilent 35670A
Ratio
NA
NA
NA
NA