Agilent 35670A

To set up the ICP supply test

Performance Test and Operation Verification
This test verifies that the Agilent 35670A meets its input specification for ICP
signal conditioning. In this test, a digital multimeter directly measures the open
circuit voltage of each channel. The digital multimeter measures the current souce
of each channel by measuring the voltage across a 50
The digital multimeter is set to the 100 V range to measure open circuit voltage and
set to the 1 V range to measure the current source.
1
2 ch
2
2 ch
Verifying Specifications
To set up the ICP supply test
feedthrough termination.
3-39