To Run The Program In Semiautomated Mode - Agilent Technologies 35670A Installation Manual

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Verifying Specifications

To run the program in semiautomated mode

To run the program in semiautomated mode
You must have an GPIB printer connected to your system to run the program in
semiautomated mode. If you do not have a printer, see ''To run the program without a
printer'' later in this chapter.
1
Press the following keys and when the program prompts you, type in the information
for the title page of the test record and press [
[
]
TITLE PAGE
[
]
TEST FACILITY
[
FACILITY ADDRESS
[
]
TESTED BY
[
]
REPORT NUMBER
[
]
CUSTOMER
[
]
MORE
[
]
TEMP
[
]
HUMIDITY
[
LINE FREQUENCY
[
]
RETURN
2
Press the following keys and when the program prompts you, type in the equipment
configuration information:
[
]
EQUIP CONFIG
[
]
AC CALIBRATO
[
]
SYNTH. 1
[
] (If needed)
SYNTH. 2
[
LOW-D OSCILLATO
[
]
MULTIMETER
[
]
RETURN
The GPIB address is 100 × (interface select code) + (primary address). The interface
select code for the test equipment and printer is 7 (for example, if the primary address
is 8, the GPIB address is 708).
When entering the calibration due date, only four characters are displayed on the
screen. However, you can enter up to nine characters and they will be printed.
3-8
]
]
] (If needed)
]:
ENTER
Agilent 35670A