Specifications apply in histogram/time mode, unfiltered time display
DC amplitude accuracy
Rise time of -1 V to 0 V test pulse
Settling time of -1 V to 0 V test pulse
Pulse aberrations (peak overshoot)
of -1 V to 0 V test pulse
Peak aberration relative to the mode-to-mode
difference (most common values)
Sampling period
1 channel mode
2 channel mode
4 channel mode (option AY6 only)

Trigger

Trigger modes
Maximum trigger delay
Post trigger
Pre trigger
No two channels can be further than ±7168
samples from each other.
External trigger maximum input
External trigger range
Low range
High range
External trigger resolution
Low range
High range
±5.0 % fs
<11.4 µs
<16 ms to 1%
<3 %
3.815 µs (1/262144 Hz) to 2 s in 2 × steps
7.629 µs (1/131072 Hz) to 4 s in 2 × steps
15.26 µs (1/65536 Hz) to 8 s in 2 × steps
Internal trigger
External trigger
Source trigger
GPIB trigger
8191 seconds
8191 sample periods
±42 Vpk
-2 V to +2 V
-10 V to +10 V
15.7 mV
78 mV