To Set Up The Input Resistance Test - Agilent Technologies 35670A Installation Manual

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Agilent 35670A

To set up the input resistance test

Performance Test only
This test verifies that the Agilent 35670A meets its input resistance specification.
In this test, a digital multimeter directly measures the input resistance of each
channel. The digital multimeter is set to the 1 M
1
2 ch
2
2 ch
Verifying Specifications
To set up the input resistance test
range.