Agilent Technologies 35670A Installation Manual page 125

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Agilent 35670A
Serial Number:_______________________Report Number:____________________
Test Date:___/___/___
ICP Supply
Measurement
Ch 1 Open Circuit Voltage
Ch 2 Open Circuit Voltage
Ch 1 Current
Ch 2 Current
Source Amplitude Accuracy
Measurement
1 kHz, 0.1 Vpk
1 kHz, 3.0 Vpk
1 kHz, 5.0 Vpk
Source Output Resistance
Measurement
Resistance
Verifying Specifications
Performance Test Record - Two Channel
Lower
Upper Limit Measured Value Pass/Fail
Limit
26 Vdc
32 Vdc
26 Vdc
32 Vdc
2.75 mA
5.75 mA
2.75 mA
5.75 mA
Lower
Upper Limit
Limit
(%)
(%)
-4
4
-4
4
-4
4
Lower
Upper Limit
Limit
(ohm)
5
13 of 14
Vdc
Vdc
mA
mA
Measured Value
Pass/Fail
(%)
Measured Value
Pass/Fail
(ohm)

Advertisement