Agilent Technologies 35670A Installation Manual page 38

Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

Warning
Preparing the Analyzer for Use
This chapter contains instructions for inspecting and installing the
Agilent 35670A Dynamic Signal Analyzer. This chapter also includes
instructions for cleaning the screen, transporting and storing the analyzer.
DC Power Requirements
The analyzer can operate from a dc power source supplying a true range of
10.8 to 30.8 Vdc. With all options installed, power consumption is less than
200 VA. The following table shows typical current requirements at different
operating voltages for the standard two-channel analyzer and for the optional
four-channel analyzer.
Operating
Voltage
Standard 2 channel Agilent 35670A
12 Vdc
24 Vdc
AC Power Requirements
The analyzer can operate from a 47 to 440 Hz, single-phase, ac power source
supplying 90 to 264 Vrms. With all options installed, power consumption is
less than 350 VA.
Only a qualified service person, aware of the hazards involved, should measure
the line voltage.
2-2
Typical Current
Optional 4 channel Agilent 35670A
8.0 amps
4.0 amps
11.0 amps
5.5 amps