Verifying Specifications
To set up the intermodulation distortion test
To set up the intermodulation distortion test
Performance Test only
This test verifies that the Agilent 35670A meets its FFT dynamic range
specification for intermodulation distortion. In this test, two signals are combined
to provide a composite signal to all channels. The intermodulation products are
found at the sum (F1 + F2) and difference (F1 - 2F2) frequencies. The analyzer
measures the amplitude of each intermodulation product.
1
2 ch
Using an HP 3326A
1
4 ch
Using an HP 3326A
3-28
Agilent 35670A