Verifying Specifications

To run the program without a printer

To run the program without a printer
Use this procedure if you do not have an GPIB printer connected to yout system.
1
Write in the information needed on the title page of the selected test record.
The test records are located near the back of this chapter and may be copied without
written permission of Agilent Technologies.
2
Press the following keys and when the program prompts you, type in the model
number and GPIB address:
[
]
EQUIP CONFIG
[
]
AC CALIBRATO
[
]
SYNTH. 1
[
] (If needed)
SYNTH. 2
[
LOW-D OSCILLATO
[
]
MULTIMETER
[
]
RETURN
The GPIB addresses equals 100 × (interface select code) + (primary address). The
interface select code for the test equipment is 7 (for example, if the primary address is
8, the GPIB address is 708).
3
Press the following keys:
[
]
TEST CONFIG
[
]
PROCEDURE
[
] or [
OP_VERIFY
PERFORMAN
[
]
STOP AFTER
[
EACH MEASUREMENT
[
]
RETURN
4
Press the following keys to start the test:
[
]
START TESTING
[
START BEGINNING
3-10
] (If needed)
]
]
]
Agilent 35670A