Calibrating The Ises; Storing The Reference Electrode - Siemens ADVIA 1800 Operator's Manual

Chemistry system
Table of Contents

Advertisement

IMPORTANT
If clogging occurs, the most probable cause is that the flow path is clogged inside the
electrode. Remove the Na and K electrodes, and check them by transmitted light to
see whether the flow path is clogged or not. You cannot do this for the Cl electrode
because of its construction. When in doubt, even if you cannot find a problem, try
replacing the electrode.
7. Mount the stainless steel cover on the top of the ISE unit by sliding it inside and
fasten the screw retaining the cover.
NOTE:
fastening the screw, verify that the cover is not caught in the groove and is not loose.
8. Reinstall the cover and tighten the screws.
9. Replace the DPP shield and secure it in place with the Phillips screw.
10. At the ISE Operation window, in the Initialize area, select
NOTE:
11. After initialization is complete, select

Calibrating the ISEs

1. At the Operation Panel, select
2. At the Menu Panel, select
3. At the ISE Operation window, in the Calibration area, select
4. When prompted, select
5. If the calibration fails, repeat calibration again and if data continues to be unstable,
perform an electrode wash.
NOTE:
calibration is achieved.
6. At the ISE Operation window, select the
electrode information.

Storing the reference electrode

1. Remove the reference electrode from the ISE module.
2. Rinse the electrode with deionized water.
3. Place it into an appropriate container.
4. Cover the reference electrode with reference electrode filling solution.
5. Cover the container and store at 0° - 40°C (32° - 104°F).
6. When ready to use, rinse the electrode with deionized water.
NOTE:
temperature before use.
176
When sliding it, be careful not to scratch the tubes or dilution bowl. When
The ISE wash is automatically turned on.
Initialize
Maint
to execute calibration.
Yes
The electrodes may have to stabilize on the system before a successful
If the electrode is stored cold, allow time for it to equilibrate to room
, then select
Exit
Yes
.
., then select
ISE Operation
Electrode Info
ADVIA 1800 Operator's Guide
Execute
, then select
Yes
.
.
.
Execute
button and enter the new
.