Siemens ADVIA 1800 Operator's Manual page 210

Chemistry system
Table of Contents

Advertisement

Item
Optional universal rack handler
dimensions
Weight
Environment
Electrical requirements
Cooling/ventilation requirements
210
Description
95.25(h) x 72.5(w) x 103.0(d) cm, (37.5 x 28.5 x 40.55 in.)
80.7 kg (178 lb)
• The system, with or without the rack handler, conforms to
Installation/Overvoltage Category II.
• Space around the system must be sufficient to allow for ventilation.
There must be a minimum of 19.7 inches (500 mm) behind the system,
32 inches (800 mm) on each side, and 42 inches (1000 mm) in front of
the system.
• The system should not be exposed to direct sunlight.
• Environment should be free of corrosive, flammable, or anesthetic gases
(the analyzer is not built to be explosion-proof); significant vibration;
and electrical disturbances, such as electromagnetic and electrostatic
induction.
• The system should be placed on a level floor (1/200 gradient or less)
that is capable of supporting a load of 560 kg (1234 lb).
• The average acoustic noise output from the analyzer is <70 dba with the
top cover open.
• A 3-kVA power source, single-phase, 2-pole, 3-wire configuration with
Class III grounding
• The following input voltages can be tapped: 200, 220, 230, or 240 VAC
at 50/60 Hz.
• Main supply voltage fluctuations not to exceed ± 10 percent of the
nominal voltage. Maximum current draw at in-rush is 15 amps at 200
VAC or 13 amps at 240 VAC.
• For the optional universal rack handler, the requirements are 110 Vac
50/60 Hz, 0.6 A.
• Facility Switch Box: Should contain a circuit breaker and knife switch
with one of the following ratings:
200 volts 15 amps, 220 volts 14 amps, 230 volts 13 amps, and 240 volts
13 amps
• Switch box should be located less than 5 m (15 feet) from the system
and easily accessible.
NOTE: Mark near the circuit breaker and knife switch that the switch
box is a disconnecting device for the equipment.
• Facility Supply Wiring: UL listed cord for external wiring in US and
Canada. (Must be CSA certified in Canada.)
Rated 300 volts or more, and 70°C or more
Size AWG # 14 or more (OD 13 to 18 mm), 3-wire configuration that
has a protective ground wire with covering material colored with
green and yellow stripe.
Circuit breaker - 15 or 20 amps.
• Ventilation sufficient to maintain +18°C to +30°C (+64° to +86°F)
operating temperature
• The maximum temperature change the system can accommodate is
2°C/hour.