Siemens ADVIA 1800 Operator's Manual page 167

Chemistry system
Table of Contents

Advertisement

1. Set the analyzer to Standby.
2. Unscrew the filter cap (
tube with the filter.
3. Disconnect the RRV bath oil bottle level sensor connector (
connector counter-clockwise and pull it out.
4. Remove the RRV bath oil bottle (
5. Refill the bottle with RRV (reaction) bath oil.
6. Replace the bottle on the shelf in the cabinet.
7. Connect the RRV bath oil bottle level sensor connector (
in and turning it clockwise.
8. Insert the filter and tube, then fasten the cap.
NOTE:
9. Return the analyzer to the READY mode.
Maintenance
CAUTION
Do not attempt to clean the RRV bath oil bottle (
and water do not mix.
) at the front of the RRV bath oil bottle (
3
CAUTION
Make a note of the bottle position on the shelf to avoid mixing up the fluid
bottles.
CAUTION
Ensure that level-sensor connector (
Make sure that the filter holder is located at the bottom of the bottle.
1
1
).
) does not get wet to avoid damaging it.
2
2
) with water; RRV bath oil
), then pull up the
1
), then turn the
2
) by pushing the connector
167