Performing The Shutdown Wash - Siemens ADVIA 1800 Operator's Manual

Chemistry system
Table of Contents

Advertisement

NOTE:
consecutive hours daily.
The daily startup wash rinses the probe lines, reaction cuvettes and dilution cuvettes.
1. At the Operation Panel, select
2. Ensure the 10-mL tube at CTT (
TIP
By choosing CTT position #50 for WASH 2 and CTT position #51 for WASH 3, you
only need to refill the CTT container once.
3. Ensure the container at RTT1 (
NOTE
on each of the trays, but you must change the entries for the alternate positions in the
appropriate fields on the WASH Set window.
4. At the WASH Set window, define the WASH3 container positions as follows:
a. Select
b. Select
c. Type
d. Type
5. Select

Performing the shutdown wash

Materials required:
• 10% solution of Cuvette Wash
Solution (REF 00195330, PN B01-
4178-01)
• Deionized water
• ISE Detergent
(REF 01307361,
PN B01-4174-01)
Time: 38 minutes
Analyzer mode: READY
The daily shutdown wash uses a detergent to clean the probe lines, reaction and dilution
cuvettes, and ISE components.
NOTE:
hours per day are advised to perform the Weekly wash procedure in place of this
Shutdown wash procedure.
For location of washes on the CTT and RTT trays, refer to Figure 5-12.
1. At the Operation Panel, select
2. Ensure the 10-mL tube at CTT (
Wash Solution and that the cup at CTT (
128
This procedure is only necessary if the analyzer is inactive for more than 4
: At your laboratory's discretion, you may use other positions for the washes
.
WASH3
for Cycles.
1
in the CTT cup position 1st time field.
51
in the RTT1 and RTT2 bottle position 1st time field.
56
Execute
.
Laboratories running only urine samples or those running the system more than 8
.
Wash
) position #51 contains DI water.
1
2
) and RTT2 (
3
) position #56 contains DI water.
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
Wash
.
1
) position #49 contains a 10% solution of Cuvette
1
) position #15 contains ISE Detergent.
ADVIA 1800 Operator's Guide