Item
Minimum sampling volume
Special dilution
Barcode capability
Dilution
Dilution tray (DTT)
Number of cuvettes
Maximum sample volume
Dilution cuvette dead volume
Dilution ratio
Reaction cuvette material
Reagent
Dispensing system
Trays
Container capacity
Refrigerator
Reagent volume/item
Dilution reagent volume
Reagent inventory
Barcode
Reaction
Reaction tray (RRV)
Number of cuvettes
RRV cuvette material
Reaction liquid volume
Stirring system
Reaction times
Reaction temperature
208
Description
1 µL (0.1 µL increments)
Diluted sample can be rediluted directly from tray.
I
2 of 5, Codabar, Code 39, Code 128
Turntable system
120 (6 sets of 20 cuvettes)
300 µL
35 µL
From 0 to 1:5625
Plastic
2-reagent capability, 2-probe system
Two trays, each holding 56 containers
Multiple reagent pack loads – 5 reagents per method with automatic
rollover upon depletion
Reagents on each tray are cooled to between
6 °C and 14 °C
20, 40, or 70 mL
All method reagents
15 to 150 µL (0.1 µL increments)
15 to 150 µL (0.1 µL increments)
Computed by the liquid-level sensing
Interleaved 2 of 5
Turntable system
221 (13 sets of 17 cuvettes)
Plastic
80 to 300 µL; minimum read-volume 80 µL
Rotation and reciprocation
Strong and weak stirring options are available.
Stirring immediately after additions of samples and reagents (S, R1, R2)
Mixer 1 is used for R1; Mixer 2 is used for R2.
3, 4, 5, 10 minutes
Extended reaction times: 15 & 21 minutes
37°C
Temperature regulation: ±0.1°C
ADVIA 1800 Operator's Guide