Setup windows

NOTE:
under the Setup button.

System Specification Settings window*

Use this window to enter settings related to the system configuration.
You can enter or change settings for the areas listed below:
• Settings that take effect after the next New Start
basic system composition
basic system operation
• Settings that take effect after startup operation from READY state
sample containers
reagent bottles
• Serum indices specifications
The settings below are related to serum information items (serum indices), which are
involved in creating workorders. If the serum indices for each sample are not selected
in the Order Entry window, select one of these settings:
Request. item range
Compulsory item analysis
Not handled
• Set the system monitor
• Set system parameters
• Basic system composition
These settings allow you to customize general components of the system:
ISE
Sample bar code
Concentrated waste
• Basic system operation
These settings are used for retesting of samples
Auto. retest
Manual retest
• Sample containers
The settings below specify the sample containers used by the system. You can specify
up to 9 container types.
System Overview
You must log on as a supervisor, tech_manager, or service to access window
CAUTION
Normal operation and the volume of sample aspirated depends on the
specifications you enter here. Incorrect settings may damage the probes and
nozzles.
Sample delivery
On-line
Reagent bar code
61